H-Index
17
Scimago Lab
powered by Scopus
eISSN: 1941-5923
call: +1.631.629.4328
Mon-Fri 10 am - 2 pm EST

Logo

MSMbanner
Medical Science Monitor Basic Research

Annals
ISI-Home

“Mind the Gap”: An 85-Year-Old Man with Severe Tricuspid Valve Regurgitation Who Underwent Percutaneous Edge-to-Edge Valve Leaflet Plication Using the New and Advanced MitraClip XTR System

Unusual clinical course, Challenging differential diagnosis, Unusual or unexpected effect of treatment, Diagnostic / therapeutic accidents, Educational Purpose (only if useful for a systematic review or synthesis)

Karin Anne Lydia Mueller, Rezo Jorbenadze, Meinrad Gawaz, Juergen Schreieck, Peter Seizer

Germany Department of Cardiology and Angiology, University Hospital, Eberhard Karls University Tuebingen, Tuebingen, Germany

Am J Case Rep 2021; 22:e928089

DOI: 10.12659/AJCR.928089

Available online: 2020-12-02

Published: 2021-01-18


#928089

BACKGROUND: Severe tricuspid valve regurgitation (TR) is associated with high cardiovascular mortality. Safe and feasible interventional approaches to treat severe TR are of clinical relevance. The MitraClip is a device that has been approved by the US Food and Drug Administration (FDA) for the repair of mitral valve lesions. Percutaneous femoral venous access with fluoroscopic and echocardiographic guidance is used to deliver a cobalt-chromium clip to secure the mitral valve leaflets. We report on an 85-year-old man with tricuspid valve regurgitation who underwent percutaneous edge-to-edge tricuspid valve leaflet plication with the new, advanced MitraClip XTR System.
CASE REPORT: An 85-year-old man with severe TR due to annulus dilation of the right ventricle and short septal leaflet presented repeatedly at our hospital with severe right heart failure symptoms. Transesophageal echocardiography revealed severe TR with a large coaptation gap size of 10.6 mm. Percutaneous edge-to-edge valve repair with the new-generation MitraClip System XTR with wider clip arms could overcome the large coaptation gap. We achieved a strong reduction of TR after deploying 2 MitraClips XTR. The patient recovered quickly and has not been admitted to hospital due to heart failure symptoms since the intervention for more than 6 months.
CONCLUSIONS: Previous studies have shown the safety and effectiveness of the MitraClip device and supported FDA approval for tricuspid valve repair. This report of a patient with complex tricuspid regurgitation demonstrated the feasible use of the new MitraClip XTR System, which improved edge-to-edge tricuspid valve repair due to its increased span and improved grip.

Keywords: Administration, Cutaneous, Cardiac Valve Annuloplasty, Heart Valves, Mortality, Tricuspid Valve Insufficiency