H-Index
14
Scimago Lab
powered by Scopus
eISSN: 1941-5923
call: +1.631.629.4328
Mon-Fri 10 am - 2 pm EST

Logo


Get your full text copy in PDF

Ultrasound techniques to measure the optic nerve sheath: is a specialized probe necessary?

Sachita Shah, Heidi Kimberly, Keith Marill, Vicki E. Noble

Med Sci Monit 2009; 15(5): MT63-68

ID: 869629


Background: Ultrasound measurement of the optic nerve sheath diameter (ONSD) has been shown to correlate with intracranial pressure (ICP). This study looked at intra-observer and inter-observer reliability among three emergency physicians (EPs) using three different imaging protocols.
Material and Method: Ocular ultrasound was performed in 20 healthy volunteers. Each physician obtained 3 measurements of the right eye ONSD with both the 10-5 MHz and 13-6 MHz probes using the axial imaging technique (experiment 1). The ONSD was then measured with the 10-5 MHz probe using the axial technique and the intracavitary 8-5 MHz probe using the coronal technique (experiment 2).
Results: In experiment 1, larger measurements were obtained with the standard 10-5 MHz probe with an adjusted difference of 0.21 mm (p<0.001). The standard deviation between intra-observer measurements for the 10-5 MHz probe was slightly but not significantly larger with an adjusted difference of 0.039 (p=0.10). In experiment 2, larger measurements were obtained with the standard 10-5 MHz probe with an adjusted difference of 0.91 mm (p<0.001). The standard deviation between intra-observer measurements was greater for the 10-5 MHz probe with an adjusted difference of 0.06 (p=0.03). The inter-observer measurements in experiment 2 had significantly larger variance with the 10-5 MHz probe (p=0.001).
Conclusions: These data suggest comparable precision and inter-observer reliability can be achieved. However, larger values when using the 10-5 MHz probe suggest different cut-off thresholds for normal upper limits. The coronal imaging technique demonstrated less inter-rater variance and may be a better technique when such small measurement differences are important.

This paper has been published under Creative Common Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0) allowing to download articles and share them with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially.